Servicing
Instruction Manual FireGuard 2
13542E/5
63
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
Only carry out this step for installation
in suspended ceilings.
Loosen and swing away the two fastening
clips (circles). Lift the photometer out of the
measuring position.
2.
Remove and clean the contamination protec-
tion (circles).
3.
Remove the measuring cell housing
(Figure 31, pos. 2) by loosening the screws on
both sides of the electronic component.
If present, The sample heater can reach
temperatures of up to 80 °C!