Servicing
Instruction Manual FireGuard 2
13542E/5
61
9
Servicing
CAUTION!
Damage to the instrument due to servicing duties that have not been carried out
or have been carried out improperly.
If servicing duties are not carried out according to the servicing schedule or non-original
SIGRIST spare parts are used, this can lead to damage to the instrument or measuring er-
rors.
In this case, SIGRIST-PHOTOMETER AG accepts no warranty claims made by the customer
and is not responsible for any subsequent costs. To avoid this situation, please adhere to
the following steps:
▪
Carry out servicing duties according to the servicing schedule (Section 9.1).
▪
When carrying out servicing duties, use original SIGRIST spare parts according to the
spare parts list (Section 15). The use of third-party spare parts requires the written ap-
proval of SIGRIST-PHOTOMETER AG.
▪
If the instruments are subjected to heavy use or exposed to severe environmental con-
ditions, servicing duties must be carried out more often as required. Wear parts in par-
ticular must be replaced more often depending on the circumstances.
9.1
Servicing schedule
WARNING!
Alarms can be triggered during servicing duties. Therefore, evaluations of the
alarm events should be suppressed in advance.
WHEN
WHO
WHAT
PURPOSE
Annually or as
needed
Operator
Cleaning the measuring
cell
Section 9.2
Obligatory measure for main-
taining measuring accuracy.
As needed or
in the event
of an adjust-
ment fault
Operator
Cleaning the optics
Section 9.3
Obligatory measure for main-
taining measuring accuracy.
Annually
Operator
Recalibration of the pho-
tometer
Section 9.4
Obligatory measure for main-
taining measuring accuracy.
Every 10 years
or as needed
Operator
Replacing the battery in
the SICON-C
Section 9.5
Obligatory measure for main-
taining functional efficiency.
Table 1: Servicing schedule