Centrifugation with low capacity
• Install the tubes axial symmetrically so that the buckets and their inserts
are loaded evenly.
• It is not permissible to load angle rotors on only one axis.
correct
incorrect
Fig. 10: Permissible and impermissible loading of an angle rotor and a swing-out rotor
(example illustrations)
Pos : 113 /010 Uni vers almodul e/ Leerz eile @ 0\ mod_1202116244500_0.doc x @ 114 @ @ 1
Pos : 114 /100 Sig ma/100 BA Z entrif ugen Sig ma (St andardmodul e)/060 Betrieb_Spi nc ontr ol L+S060-0020- 0020-0040 Adapter - nur L+ S @ 27\ mod_1405319210191_68. doc x @ 192563 @ 4 @ 1
→
6.2.2.4
Adapters
In order to ensure easy handling, even if vessels of various sizes are used,
carrier systems were developed.
• Load the opposite adapters with the same number of vessels and with
the same weights in order to avoid imbalance.
• If all of the compartments of a carrier are not used, the buckets must be
loaded evenly. Loading the edges of a bucket only is not permissible.