Refrigerated Centrifuge Sigma 6-16KS
6 Using the centrifuge
30 / 91
Version 06/2012, Rev. 1.11 of 06/04/2016 • sb
Translation of the original operating manual, part no. 0702202
Pos : 112 /100 Sig ma/100 BA Z entrif ugen Sig ma (St andardmodul e)/060 Betrieb_Spi nc ontr ol L+S060-0020- 0020-0032 Ei ns etz en von Zubehör 4+6plät zig 4- 16S bis 6-16KS @ 27\ mod_1405319207709_68.doc x @ 192535 @ 4 @ 1
→
6.2.2.3
Installation of accessories
• Only use inserts that are suitable for the rotor (see chapter 11.1 -
"Range of accessories ").
• All buckets of the swing-out rotor need to be installed when spinning.
• Always load the axial symmetrical inserts/buckets of the rotors with the
same accessories and fill to avoid imbalance.
Centrifugation with different tube sizes
Working with different tube sizes is possible. In this case, however, it is
very important that axial symmetric inserts are identical (see figure).
correct
incorrect
Fig. 9: Permissible and impermissible loading of a swing-out rotor with different tube sizes
(example illustrations)