Refrigerated Centrifuge Sigma 6-16KS
7 Malfunctions and error correction
58 / 91
Version 06/2012, Rev. 1.11 of 06/04/2016 • sb
Translation of the original operating manual, part no. 0702202
Pos : 168 /100 Sig ma/110 BA Z entrif ugen Sig ma (Proj ekt e)/ 6-16KS_6- 16KHS/070 Stör ungen und F ehl ersuche070-0010-0010 Notentrieg elung @ 29\ mod_1405327259063_68.doc x @ 199493 @ 3 @ 1
7.1.1
Emergency lid release
In the event of a power failure, it is possible to manually open the centrifuge
lid.
• Switch off the mains power switch and disconnect the power cord from
the socket.
• Remove the plugs (see figure, item 1) from the opening on the left and
on the right side of the control panel, e.g. with a screwdriver.
Fig. 32: Position of the opening for the emergency lid release
• Insert the supplied hexagon socket key horizontally into the hole. The
key will be guided through a funnel-shaped tube to the shaft of the lid
lock motor.
Fig. 33: The emergency lid release key must be inserted horizontally (similar to illustration)
• Unlock the motorised lid locks as follows:
– Turn the left lid lock anti-clockwise.
– Turn the right lid lock clockwise.
• Then, reinsert the plugs.
WARNING
Do not unlock or open the lid unless the rotor is at a standstill.
If the lid is opened via the emergency lid release system during a centrifuge
run, the centrifuge will be switched off immediately and decelerate in an
unbraked manner.