4-16KR
Use as a robot-controlled Centrifuge
Version 10/2010, Rev. 1.9 of 02/07/2012
43
Translation of the original operating manual
This command is only available for Spincontrol comfort, Spincontrol
professional and Spincontrol S.
setcurve
This command sets new data for the free
programmable curves. The command is followed by the
parameter:
setcurve [curveNr], [Lin/Quad], [Int1Time], [Int1Speed],
[Int2Time], [Int2Speed],etc.
Notice this command is only possible if no free curve is running!
SIGMA> setcurve 22,0,130,112,60,148,60193,60,256,60326,60,524,60,1010
Ok
SIGMA> getcurve 22
CurveNr: 22
Interval 1: Time: 130 Speed: 112 LIN
Interval 2: Time: 60 Speed: 148
Interval 3: Time: 60 Speed: 193
Interval 4: Time: 60 Speed: 256
Interval 5: Time: 60 Speed: 326
Interval 6: Time: 60 Speed: 524
Interval 7: Time: 60 Speed: 1010
Interval 8: Time: 0 Speed: 0
Interval 9: Time: 0 Speed: 0
Interval 10: Time: 0 Speed: 0
TotalTime: 490
SIGMA>
This command is only available for Spincontrol comfort, Spincontrol
professional and Spincontrol S.
getname
Displays name of the centrifuge (given by centrifuge
menu Setup System Name).
This command is only available for Spincontrol comfort, Spincontrol
professional and Spincontrol S.
getlastrun
This command triggers output of parameters and
results of last spin in cvs-Format.
This command is only available for Spincontrol S.