4-16KR
Use as a robot-controlled Centrifuge
38
Version 10/2010, Rev. 1.9 of 02/07/2012
Translation of the original operating manual
6.3.3.2
General specifications
Interface standard:
RS232
Baud rate:
9600
Parity:
No
Data bits:
8
Stopbits:
1
Data format:
ASCII
The serial communication works without hardware- or XON/XOFF
software handshake.
6.3.3.3
Communication protocol
User commands consist of an ASCII-coded command string and - if
needed - a parameter separated by a space. The command parser
works non case sensitive.
The character received won't be echoed by the centrifuge processor
normally. You can tell the centrifuge to echo each character by
sending the
“echoon”
command. The user command and the return
string of the centrifuge will always be terminated with the characters
‘0x0A’ and ‘0x0D’ (CR and LF).
The command
“cmderror”
can be used to ensure the correct
execution of the last command.
User commands
The following categories of user commands are available:
control commands
commands to change
s
et
p
oints
(OUT_SP_n y)
commands to request
s
et
p
oints
(IN_SP_n)
commands to request
p
rocess
v
alues
(IN_PV_n)
commands to change
par
ameters
(OUT_PAR_n y)
commands to request
par
ameters
(OUT_PAR_n)
other commands