4-16KR
Using the Centrifuge
34
Version 10/2010, Rev. 1.9 of 02/07/2012
Translation of the original operating manual
6.2.3.9
Rotor selection
The inserted rotor mus be selected manually when the centrifuge is at
a standstil.
Rotor selection:
Turn the left function key until “set” appears in the display field in
front of the parameter "Rotor" (Fig. 6.15).
Select the option by pressing or turning the right function key.
"Set" will now flash.
Turn the right function key until the desired rotor number is
displayed.
Press the right function key to confirm the selected value. If this
is not done, the value will be automatically reset to the last
setting.
Fig. 6.15:
Rotor selection
6.2.4 Program Mode
The program is used to save or load certain recurrent settings of the
centrifuge. 50 different programs can be saved and called up. This is
only possible at a standstill.
Attention: Temperature profiles cannot be created!
With a program, certain sedimentation results can be repeated under
identical conditions.
A maximum of 50 programs can be stored under the numbers 1 - 50.
The rapid cooling program "RAPID_TEMP" does not occupy any
storage location and cannot be deleted. It is used to cool the
centrifuge without vessels.
6.2.4.1
Saving the Current Settings
Turn the left function key until “save” appears on the display
(Fig. 6.16).
Select the option by pressing the right function key. “Save” will
now flash.
Turn the right function key until the desired program number is
displayed. Empty storage location numbers flash.
Press the right function key to confirm the input.