11.1 Range of accessories
Part no.
Description
Max.
speed
(rpm)
Max.
gravitational
field (x g)
11030
Swing-out rotor, 4 place, for buckets no. 13031, 13033, 13034, 13035,
13037, 13038, 13099; buckets 13034, 13037 for rotors from the year 2003
onwards only
4 000
11031
Swing-out rotor 16 x 15 ml complete, no. 11030, incl. 2 sets carriers no.
13035 and 16 PS-tubes 15 ml no. 15020,
Radius max.13.3 cm, radius min. 4.9 cm
4 000
2 379
11032
Swing-out rotor 4 x 100 ml complete, no. 11030, incl. 2 sets buckets no.
13099 and 4 glass tubes 100 ml no. 15100,
Radius max.13.4 cm, radius min.5.0 cm
4 000
2 397
11035
Swing-out rotor 8 x 15 ml complete, no. 11036, incl. 1 set carriers no. 13035
and 8 PS-tubes 15 ml no. 15020
Radius max. 13.3 cm, radius min. 4.9 cm
4 000
2 379
11036
Swing-out rotor, 2 place, for carriers no. 13031, 13033, 13034, 13035,
13038, 13099
4 000
11070
Swing-out rotor, 4 place, for bucket no. 13296 and 13299, Software version
22 or higher
4 000
11121
Swing-out rotor for microtiter plates, incl. 1 set carriers no. 13223, max.
height of plates 50 mm, max. load per bucket approx. 345 gram,
Radius max. 11.8 cm, radius min. 4.8 cm
4 000
2 111
12011
Angle rotor 6 x 15 ml complete, incl. 3 sets buckets no. 13011, for tubes Ø
17 x up to 130 mm, e.g. no. 15015, 15020, 15023, 15024, culture tubes 15
ml no. 15115 and for 6 reaction vials 1.5/2.0 ml no. 15008, 15040,
Radius max. 9.7 cm, radius min. 3.6 cm, angle 35°
4 000
1 735
12012
Angle rotor 8 x 15 ml complete, incl. 4 sets buckets no. 13011, for tubes Ø
17 x up to 115 mm, e.g. no. 15015, 15020, 15023, 15024 and for 8 reaction
vials 1.5/2.0 ml no. 15008, 15040
Radius max. 9.7 cm, radius min. 3.6 cm, angle 35°
4 000
1735
12061
Angle rotor 30 x 15 ml complete, incl. 15 sets buckets no. 13011, for tubes
Ø 17 x up to 120 mm, e.g. no. 15015, 15020, 15023, 15024, Monovettes
and culture tubes 15 ml no. 15115
Radius max. 13.9 cm, radius min. 5.8 cm, angle 33°
Warning!
This rotor may heat up to more than +15°C during long runs of more
than 15 minutes or repeated runs at short intervals!
4 000
2 486
12062
Angle rotor 20 x 15 ml complete, incl. 10 sets buckets no. 13011, for tubes
Ø 17 x up to 120 mm, e.g. no. 15015, 15020, 15023, 15024, Monovettes
and culture tubes 15 ml no. 15115,
Radius max. 13.9 cm, angle 33°
Warning!
This rotor may heat up to more than +15°C during long runs of more
than 15 minutes or repeated runs at short intervals!
4 000
2 486
13031
Bucket for 1 culture tube e.g. 15151 incl. polysulfone, cap no. 17151,
1 set = 2 pcs., suitable for 11030, 11036,
Radius max.14.5 cm, radius min. 4.3 cm
4 000
2 594