![Sigma 2-6 Operating Manual Download Page 42](http://html1.mh-extra.com/html/sigma/2-6/2-6_operating-manual_1267893042.webp)
Laboratory Centrifuge Sigma 2-6
8 Maintenance and service
42
Version 04/2007, Rev. 2.10 of 06/11/2015 • sb
Translation of the original operating manual, part no. 07044
Pos: 162 /10 0 Sigm a/10 0 BA Z entrif uge n Sigm a (Sta nda rd mod ule)/ 080 W art ung und Inst andh altun g08 0-0 010 -00 20 Zu beh ör (au ßer 8K) @ 28\ mo d_1 4053 206 182 47_ 68.d ocx @ 194 845 @ 3 @ 1
8.1.2 Accessories
CAUTION
For the care of the accessories, special safety measures must be
considered as these are measures that will ensure operational safety at
the same time!
• Immediately rinse off the rotor, buckets, or accessories under running
water if they have come into contact with any liquids that may cause
corrosion. Use a brush for test tubes in order to clean the bores of angle
rotors. Turn the rotor upside down and allow it to dry completely.
• Clean the accessories outside the centrifuge once a week or preferably
after each use. Adapters should be removed, cleaned and dried.
CAUTION
Do not clean the accessories in a dishwasher!
Cleaning in a dishwasher removes the anodised coating; the result is
cracking in areas that are subject to stress.
• If the rotors or accessories have been contaminated with toxic,
radioactive, or pathogenic substances, clean them immediately with a
suitable decontamination agent (depending on the type of
contamination).
Take suitable precautions for your own safety if there is
a risk of toxic, radioactive, or pathogenic contamination.
• Dry the accessories with a soft cloth or in a drying chamber at
approx. 50°C.
Pos: 163 /01 0 Univ ersal mod ule/L eerz eile @ 0\m od_ 120 211 624 450 0_0. docx @ 11 4 @ @ 1
Pos: 164 /10 0 Sigm a/10 0 BA Z entrif uge n Sigm a (Sta nda rd mod ule)/ 080 W art ung und Inst andh altun g08 0-0 010 -00 20- 001 0 Ku nststo ffzub ehö r @ 28\ mod _140 532 061 960 2_6 8.doc x @ 1 948 59 @ 4 @ 1
8.1.2.1 Plastic accessories
The chemical resistance of plastic decreases with rising temperatures (see
chapter 11.4 - "Resistance data").
• If solvents, acids, or alkaline solutions have been used, clean the plastic
accessories thoroughly.
WARNING
Plastic accessories must not be greased!
Pos: 165 /01 0 Univ ersal mod ule/Seit enwec hsel @ 0\m od_ 120 211 624 4312 _0. docx @ 10 5 @ @ 1