Laboratory Centrifuge Sigma 2-6
6 Using the centrifuge
Version 04/2007, Rev. 2.10 of 06/11/2015 • sb
33
Translation of the original operating manual, part no. 07044
Pos: 125 /10 0 Sigm a/10 0 BA Z entrif uge n Sigm a (Sta nda rd mod ule)/ 062 B etrie b_Spinc ont rol Easy +Unive rsal/Spi ncont rol Easy 062 -00 30- 002 0-0 050 Dr ehza hl/RZB @ 28\ mod _14 053 204 738 39_ 68.d ocx @ 194 243 @ 4 @ 1
6.3.2.6 Speed / relative centrifugal force (RCF)
The set speed of the centrifuge is displayed in the upper area of the
Speed/RCF field. The actual value is displayed right below. The speed is
stated as the number of revolutions per minute (rpm) and the RCF values
as a multiple of the gravitational acceleration (x g). The values are
interdependent (see chapter 2.2.2.1 - "Speed, radius, and relative
centrifugal force"). The maximum speed/RCF values depend on the rotor
that is used.
The parameters speed and RCF can be changed during the centrifugation.
Fig. 13: Setting the speed value or the RCF value
Changing the speed increments
By default, the speed is changed in intervals of 100 rpm.
To change the speed in intervals of 1 rpm:
• Keep the stop key pressed while setting the desired speed with the right
function knob.
Changing the RCF increments
By default, the RCF value is changed in intervals of 10 x g.
To change the RCF value in intervals of 1 x g:
• Keep the stop key pressed while setting the desired RCF value with the
right function knob.