Laboratory Centrifuge Sigma 2-6
5 Set-up and connection
26
Version 04/2007, Rev. 2.10 of 06/11/2015 • sb
Translation of the original operating manual, part no. 07044
Pos: 86 / 100 Sigma /100 BA Ze ntrifu gen Sigma (Stan da rdm odule )/0 50 Aufs tellun g u nd Ansc hluss0 50 Au fstellu ng u nd Ansc hluss === == == === == == == === == == === == == === = @ 27\ mod _14 053 1917 307 1_6 8.do cx @ 1923 81 @ 1 @ 1
5 Set-up and connection
Pos: 87 / 100 Sigma /100 BA Ze ntrifu gen Sigma (Stan da rdm odule )/0 50 Aufs tellun g u nd Ansc hluss0 50- 001 0 Aufs tellor t @ 2 7\m od_ 140 531 917 4276 _68 .docx @ 1 923 95 @ 2 @ 1
5.1 Installation site
Operate the centrifuge only in closed and dry rooms.
All the energy supplied to the centrifuge is converted into heat and emitted
to the ambient air.
• Ensure sufficient ventilation.
• Keep a safety distance of at least 30 cm around the centrifuge so that
the vents in the centrifuge remain fully effective.
• Do not subject the centrifuge to thermal stress, e.g. by positioning it
near heat generators.
• Avoid direct sunlight (UV radiation).
• The table must be stable and have a solid, even surface.
• Attention: During transport from cold to warmer places, condensational
water will collect inside the centrifuge. It is important to allow sufficient
time for drying (min. 24 h) before the centrifuge can be used again.
Pos: 88 / 010 Unive rsalm odul e/Le erzeil e @ 0\mo d_1 202 116 244 500 _0.d ocx @ 114 @ @ 1
Pos: 89 / 100 Sigma /100 BA Ze ntrifu gen Sigma (Stan da rdm odule )/0 50 Aufs tellun g u nd Ansc hluss0 50- 002 0 En ergie vers org ung --- -- --- -- --- --- -- --- -- --- -- --- --- -- --- -- --- --- -- --- -- @ 27 \mo d_1 405 319 1756 30_ 68. docx @ 192 409 @ 2 @ 1
5.2 Power supply
Pos: 90 / 100 Sigma /100 BA Ze ntrifu gen Sigma (Stan da rdm odule )/0 50 Aufs tellun g u nd Ansc hluss/0 50 -002 0-0 010 Anschl ussa rt (Kalt ger ätes tecke r) @ 32\ mo d_14 175 996 706 21_ 68.d ocx @ 222 346 @ 3 @ 1
5.2.1 Type of connection
DANGER
The operating voltage on the name plate must correspond to the local
supply voltage!
Sigma centrifuges are units of protection class I. The centrifuges of this
model series have a three-wire power cord with an IEC C13 connector.
Pos: 91 / 010 Unive rsalm odul e/Le erzeil e @ 0\mo d_1 202 116 244 500 _0.d ocx @ 114 @ @ 1
Pos: 92 / 100 Sigma /100 BA Ze ntrifu gen Sigma (Stan da rdm odule )/0 50 Aufs tellun g u nd Ansc hluss/0 50 -002 0-0 020 Siche run gen baus eits (au ßer 3- 30KS) @ 7\m od_ 130 6312 302 397 _68 .docx @ 43 513 @ 3 @ 1
5.2.2 Customer-provided fuses
Typically, the centrifuge must be protected with 16 Amp L or B fuses that
are to be provided by the customer.