SIGLENT
SPD3000 User Manual 15
2.5 CH1/CH2 Parallel mode
Instruction
In the parallel mode, CH1 and CH2 are linked internally into one
channel which is controlled by CH1. Its output current value is
twice as much as the single channel.
Output rating 0~30V/0~6A
Operation steps:
1. Press the “PAR” button to start Parallel mode, and the indicator light turns
bright immediately
;
2. Connect the load to the positive and negative terminal of CH1
;
3. Turn on CH1, and then set the output voltage/current value by revolving the
multi-function knob
;
4. Press “ON/OFF” button to open the output.
Note: You can identity the current working state “CC” or ”CV” by
referring to indicator light of CH1/CH2.(red means CV, green means
CC)
;
In parallel mode, CH2 only works in CC mode.