Page 6 of 6
7482255 Rev 01
Siemens
October 2002
Planned Maintenance Checklist
Passed
Failed N/A
Preparation
Discuss customer concerns
Clean the system
Clean the subsystems
Clean the air filter element.
Clean the track ball.
Clean and Inspect
Transducers
Safety Checks
Inspect the wheels and wheel locks.
Verify that the monitor is secure.
Perform electrical safety test:
Field Electrical Safety Test Procedure
(p/n 7482776)
G60 S Electrical Safety Test Limits
(p/n 7482784)
Field Electrical Safety Test Report
(p/n 7482768)
System Test
Calibration is successful.
Image and overlay okay.
Time and date are correct.
Functional test of controls successful.
Transducer operational check okay.
Freeze function okay.
Peripherals operate properly.
Preset data okay.
Protective ground wire operational.
Summary of Contents for SONOLINE G50
Page 4: ...Page iv 7482255 Rev 01 Siemens October 2002 G50 and G60 S Service Manual ...
Page 10: ...Page x 7482255 Rev 01 Siemens October 2002 G50 and G60 S Service Manual ...
Page 22: ...Page 12 of 12 7482255 Rev 01 Siemens October 2002 Mechanical and Electrical Specifications ...
Page 140: ...Page 8 of 8 7482255 Rev 01 Siemens October 2002 Verifying the Operation of the Software ...
Page 190: ...Page 8 of 8 7482255 Rev 01 Siemens October 2002 Concluding the Maintenance ...
Page 196: ...Page 2 of 2 7482255 Rev 01 Siemens October 2002 Figure B 1 Installation Report Form ...
Page 214: ...Page 18 of 18 7482255 Rev 01 Siemens October 2002 G60 S Spare Parts Photographs ...
Page 216: ...Page 2 of 2 7482255 Rev 01 Siemens October 2002 ...
Page 222: ...Page 6 of 6 7482255 Rev 01 Siemens October 2002 ...
Page 226: ...Page 4 of 4 7482255 Rev 01 Siemens October 2002 ...
Page 228: ...Page 2 of 6 7482255 Rev 01 Siemens October 2002 This page intentionally left blank ...
Page 234: ...Page 2 of 2 7482255 Rev 01 Siemens October 2002 ...