Page 8 of 42
7482255 Rev 01
Siemens
October 2002
Installing or Updating the DIMAQ-IP Software
1.
Back up all exam and patient data to a CD before continuing with the
DIMAQ-IP software update.
2.
On the G60 S, save the feature key for the Stress Echo option by perform-
ing the procedure "Uninstalling the Product Feature Keys" (page 8-14).
3.
Open the upper-rear panel of the system to expose the Microsoft Windows
label attached to the DIMAQ-IP.
4.
Record the 20 characters of the Microsoft Windows 2000 Product Identifi-
cation code; this license code is required to enable the new software.
5.
Close the upper-rear panel.
6.
Press the
F4
key and select the
Service
menu item.
7.
Access the Service software according to the instructions in "Service Soft-
ware Operation" (page 8-17).
8.
Select the DIMAQ-IP Maintenance button located on the Main Menu
screen (Figure 8-10) to display the Windows 2000 screen running on the
DIMAQ-IP module.
9.
Insert the CD containing the version 1.2 DIMAQ-IP software, in the CD
drive.
10.
Move the mouse pointer to the bottom of the screen to display the Windows
toolbar, and select the Start button.
11.
Select
Shutdown
→
Restart
from the menu.
12.
The DIMAQ-IP will reboot from the software CD; the display of the BIOS
and Windows boot screens will be followed by the software installation win-
dow.
13.
The blue Radisys logo will be displayed.
14.
The software installation requires approximately 5 minutes: The
R:\>
prompt appears when the installation is completed.
15.
After the
R:\>
prompt appears, remove the software CD.
16.
Operate the power switch to shut down the system: After two minutes, if the
system has not powered down, operate the circuit breaker located at the
rear of the system to de-energize it.
17.
After the system has powered down, wait 15 seconds and re-energize the
system.
!
Caution:
Failure of the DIMAQ-IP software installation or update will not nor-
mally destroy presets or service data. As a precaution, this data can be backed
up according to the procedures provided in "Backup/Restore" (page 8-36) and
at step 3. (backing up presets) on page 8-2.
Note:
The Stress Echo feature key must be reinstalled after the DIMAQ-IP soft-
ware is updated. See "Installing the Product Feature Keys" (page 8-10).
Note:
The DIMAQ-IP will boot from the CD drive if Windows is restarted with a
bootable CD installed in the CD drive.
Summary of Contents for SONOLINE G50
Page 4: ...Page iv 7482255 Rev 01 Siemens October 2002 G50 and G60 S Service Manual ...
Page 10: ...Page x 7482255 Rev 01 Siemens October 2002 G50 and G60 S Service Manual ...
Page 22: ...Page 12 of 12 7482255 Rev 01 Siemens October 2002 Mechanical and Electrical Specifications ...
Page 140: ...Page 8 of 8 7482255 Rev 01 Siemens October 2002 Verifying the Operation of the Software ...
Page 190: ...Page 8 of 8 7482255 Rev 01 Siemens October 2002 Concluding the Maintenance ...
Page 196: ...Page 2 of 2 7482255 Rev 01 Siemens October 2002 Figure B 1 Installation Report Form ...
Page 214: ...Page 18 of 18 7482255 Rev 01 Siemens October 2002 G60 S Spare Parts Photographs ...
Page 216: ...Page 2 of 2 7482255 Rev 01 Siemens October 2002 ...
Page 222: ...Page 6 of 6 7482255 Rev 01 Siemens October 2002 ...
Page 226: ...Page 4 of 4 7482255 Rev 01 Siemens October 2002 ...
Page 228: ...Page 2 of 6 7482255 Rev 01 Siemens October 2002 This page intentionally left blank ...
Page 234: ...Page 2 of 2 7482255 Rev 01 Siemens October 2002 ...