Chapter 6 - Troubleshooting
After completion of the boot sequence, the LEDs shown in Figure 6-3 indicate
the following conditions:
LED 1 illuminates only during the boot sequence. When Off, it indicates
that the boot sequence has been completed.
LED 2 illuminates after completion of the boot sequence, if the Windows-
CE and system software have been initialized.
LED 3 flashes after completion of the boot sequence, when keys, encod-
ers, or volumes are operated.
LED 4 illuminates after completion of the boot sequence to indicate that the
DSP on the DPPR has malfunctioned.
LED 5 illuminates after completion of the boot sequence, to indicate that
the DIMAQ-IP module is connected.
LED 6 flashes at one second intervals to indicate that the CPU is operating.
LED 7 illuminates when a cooling fan fails: The MO fan, two E module fans,
and two switching regulator fans are included in this circuit.
LED 8 illuminates when the temperature in the E module exceeds the over-
temperature limit of 80 °C (176 °F).
4.0 Troubleshooting the DIMAQ-IP
Use the following instructions to troubleshoot the DIMAQ-IP.
4.1 Review Switch Failure
If the DIMAQ-IP screen does not appear when the
REVIEW
control is selected,
use the following instructions to troubleshoot the failure.
1.
Verify that a patient name and ID number are entered.
2.
Verify that HDD is selected (not MOD) for Storage on the Preset Menu.
3.
Access DIMAQ-IP through the Service Menu by selecting DIMAQ-IP Main-
tenance; a blank screen indicates a DIMAQ-IP failure.
If the trackball and
SET
or
ENTER
operate properly, the DIMAQ-IP soft-
ware and Windows are running, but applications are not active. Loss of the
trackball control indicates the loss of the internal Ethernet communication.
Restart the DIMAQ-IP by rebooting Windows 2000 (
Start
→
Restart
).
Figure 6-3
Diagnostic LED Status After Boot Completed
System without DIMAQ-IP after Boot
System with DIMAQ-IP after Boot
Summary of Contents for SONOLINE G50
Page 4: ...Page iv 7482255 Rev 01 Siemens October 2002 G50 and G60 S Service Manual ...
Page 10: ...Page x 7482255 Rev 01 Siemens October 2002 G50 and G60 S Service Manual ...
Page 22: ...Page 12 of 12 7482255 Rev 01 Siemens October 2002 Mechanical and Electrical Specifications ...
Page 140: ...Page 8 of 8 7482255 Rev 01 Siemens October 2002 Verifying the Operation of the Software ...
Page 190: ...Page 8 of 8 7482255 Rev 01 Siemens October 2002 Concluding the Maintenance ...
Page 196: ...Page 2 of 2 7482255 Rev 01 Siemens October 2002 Figure B 1 Installation Report Form ...
Page 214: ...Page 18 of 18 7482255 Rev 01 Siemens October 2002 G60 S Spare Parts Photographs ...
Page 216: ...Page 2 of 2 7482255 Rev 01 Siemens October 2002 ...
Page 222: ...Page 6 of 6 7482255 Rev 01 Siemens October 2002 ...
Page 226: ...Page 4 of 4 7482255 Rev 01 Siemens October 2002 ...
Page 228: ...Page 2 of 6 7482255 Rev 01 Siemens October 2002 This page intentionally left blank ...
Page 234: ...Page 2 of 2 7482255 Rev 01 Siemens October 2002 ...