Siemens
7482255 Rev 01
Page 5 of 8
October 2002
Chapter 7 - Installing the Version 1.2 Software Update on the G60 S
15.
Select the DIMAQ-IP Maintenance button located on the Main Menu
screen to display the Windows 2000 screen running on the DIMAQ-IP mod-
ule.
16.
Insert the CD containing the version 1.2 DIMAQ-IP software, in the CD
drive.
17.
Move the mouse pointer to the bottom of the screen to display the Windows
toolbar, and select the Start button.
18.
Select
Shutdown
→
Restart
from the menu.
19.
The DIMAQ-IP will reboot from the software CD; the display of the BIOS
and Windows boot screens will be followed by the software installation win-
dow.
20.
The blue Radisys logo will be displayed.
21.
The software installation requires approximately 5 minutes: The
R:\>
prompt appears when the installation is completed.
22.
After the
R:\>
prompt appears, remove the software CD.
23.
Operate the power switch to shut down the system: After two minutes, if the
system has not powered down, operate the circuit breaker located at the
rear of the system to de-energize it.
24.
After the system has powered down, wait 15 seconds and re-energize the
system and allow it to boot.
25.
Press the
F4
key and select the
Service
menu item.
26.
Enter the service password.
27.
Select the DIMAQ-IP Maintenance button located on the Main Menu
screen to display the Windows 2000 screen running on the DIMAQ-IP mod-
ule.
28.
Enter the Windows 2000 Product Identification code and select the Ok but-
ton.
29.
Wait until the DIMAQ-IP powers down, then de-energize the system.
30.
Wait 15 seconds and re-energize the system.
31.
Allow the DIMAQ-IP to boot completely; the boot process requires approx-
imately 5 minutes.
32.
Press the Review button.
33.
Observe the DIMAQ Study Screen with the HD radio button selected; this
indicates the success of the software update.
34.
Create an image file, using any transducer, and assign a patient name and
patient identification number to it; this file will be used to verify the operation
of the CD.
35.
Export the image file to the CD to verify its operation.
36.
Verify that the exported image can be viewed by the system.
37.
Delete the exported image from the HDD.
Summary of Contents for SONOLINE G50
Page 4: ...Page iv 7482255 Rev 01 Siemens October 2002 G50 and G60 S Service Manual ...
Page 10: ...Page x 7482255 Rev 01 Siemens October 2002 G50 and G60 S Service Manual ...
Page 22: ...Page 12 of 12 7482255 Rev 01 Siemens October 2002 Mechanical and Electrical Specifications ...
Page 140: ...Page 8 of 8 7482255 Rev 01 Siemens October 2002 Verifying the Operation of the Software ...
Page 190: ...Page 8 of 8 7482255 Rev 01 Siemens October 2002 Concluding the Maintenance ...
Page 196: ...Page 2 of 2 7482255 Rev 01 Siemens October 2002 Figure B 1 Installation Report Form ...
Page 214: ...Page 18 of 18 7482255 Rev 01 Siemens October 2002 G60 S Spare Parts Photographs ...
Page 216: ...Page 2 of 2 7482255 Rev 01 Siemens October 2002 ...
Page 222: ...Page 6 of 6 7482255 Rev 01 Siemens October 2002 ...
Page 226: ...Page 4 of 4 7482255 Rev 01 Siemens October 2002 ...
Page 228: ...Page 2 of 6 7482255 Rev 01 Siemens October 2002 This page intentionally left blank ...
Page 234: ...Page 2 of 2 7482255 Rev 01 Siemens October 2002 ...