Safety notes
SMD 18101 CS / SMD 18101 CP / SMD 18101 D
2-2
Instruction Manual, 11/2005, A5E00472377A-002
Caution
Incorrect installation may result in extensive damage to property.
Installation must be carried out by specialists:
1.
When installing your medical electrical system with our products in an environment with
patients, please observe the safety requirements of
EN 60601-1-1 (IEC 601-1-1) "Specifications for the safety of medical electrical systems"
in order to prevent injury to patients and users of your systems.
2.
Use appropriate measures to ensure that the leakage currents in particular remain below
the necessary limits. Appropriate measures are:
–
Separators for signal input or signal output unit
–
Use of a safety isolating transformer
–
Use of the additional protective conductor connection
3.
Only use video cables with BNC connectors and interface cables specified by the system
manufacturer for the installation.
4.
The serial interface cable must have a female Sub-D connector at the computer end.
5.
Use line cords with contactor contact. Only insert into sockets with contactor contact.
6.
For certain applications, the video ground can be connected to the protective ground via
the additional protective conductor in the plug panel. Observe EN 60 601-1-1 (IEC 601-1-
1).
7.
Close the plug panel using the provided cover, and lock with the screw.
8.
Mounting information:
The stability of the display must be ensured following mounting of the stand/holder. The
mounting screws must only have the defined length and the defined strength.
All these requirements are satisfied when using the original stand. These requirements
must be observed when using customer-specific mounting solutions. In addition, the
country-specific regulations regarding steadiness must be observed.
9.
Note for users:
The closed connection panel must not be opened by users.
10.
Servicing information:
If housing components have to be removed for servicing, this must not be carried out in
the presence of patients, users, or other persons not involved with servicing. Displays
with a VESA mounting surface must only be connected to the power supply with the
VESA board screwed on.
11.
The following applies to installations in the USA and Canada:
Molded power supply connectors must comply with the requirements for "hospital grade
attachments" CSA Std. C22.2 No. 21 and UL 498.