Messages and Diagnostics
SIWAREX FTA
7-146
Error
No.
Data and
command errors -
description
Solution
105-
110
Plausibility error
Base parameter
105 Weighing operating
mode not allowed
or unknown
Weighing operating mode in DR 4 is not correct. Enter a correct
weighing operating mode.
106 Limit values
implausible
The default limit values in DR 4 are not plausible. Limit value 3
may only be operated as a max-limit value (On
≥
Off). Correct
limit value 3.
107
Reserved
108
Reserved
109
Reserved
110
Reserved
111-
135
Plausibility error
Weighing
parameter
111 Number tare-zero
position cycles not
allowed
The number for the tare/zero definition cycle is incorrect
(max. 100). Enter a correct value.
112 Value for control
weighing pauses
incorrect
The default for pauses between the control weighing procedures
is incorrect (max. 100). Correct the definition.
113 Empty overlapping
time too long
The emptying overlap time in DR 23 is too large. The maximum
may only correspond with the emptying time that is defined there.
114 Min.-/Max.-tare
implausible
The limits for tare in DR 23 are not plausible. Tare minimum and
maximum weight may not be less. The tare maximum weight
must be larger than the minimum weight. Correct the entry.
115 Log text No.
assignment for
automatic logging is
incorrect
The log text number in DR 23 is incorrect.
Correct the number (0...4).
116 Operating mode not
allowed or unknown
The default for the tare/zero setting mode in DR 23 is incorrect.
Correct the default (0..4).
117 Fine time too short
Fine time must be greater than Inhibition time Fine.
Correct the default in DR 23.
118 Tolerance range
implausible
The default for tolerance ranges TO1, TU1, TO2, TU2 must be
corrected. TU2 must be greater than TU1 and TO2 greater than
TO1. Check parameter definitions in DR 22.
119
Reserved
120
Reserved
121
Reserved
122
Reserved
123
Reserved
124
Reserved
125
Reserved
126
Reserved
127
Reserved
128
Reserved
129
Reserved
130
Reserved
131
Reserved