Parameter reference
SITRANS LR250 (PROFIBUS PA)
Operating Instructions, 08/2014, A5E32221386-AC
139
Event Status (3.5.2.)
Event Status allows you to assign one of the status options listed below, to any of the events
listed in Event Index (3.5.1.). This allows you to tailor a device response appropriate for your
particular process. (Event status affects Condensed status). See Condensed Status (Page 270)
for more details.
Event Status options
Good
Good: maintenance required
Good: maintenance demanded
Uncertain: maintenance demanded
* Bad: maintenance alarm
Uncertain: process related, no maintenance
Bad: process related, no maintenance
Bad: function check/local override
Good: function check
To assign a status to a particular event via the handheld programmer:
1.
Go to Enable (3.4.1.) and select Yes to enable Condensed Mode.
2.
Go to Event Index (3.5.1.) and enter the event index number corresponding to a particular
event.
3.
Go to Event Status (3.5.2.) and choose a Status option from the table above.
To assign a status to a particular event via SIMATIC PDM:
1.
Go to Level Meter > Diagnostics > Condensed Status Setup, select Yes to enable
Condensed Status Mode.
2.
Go to Level Meter > Diagnostics > Condensed Status.
3.
Select the Status line for the selected Event, then choose a Status option from the associated
pull-down menu.