Functions
7.4 Sensor characteristics
MAG 6000 I / MAG 6000 I Ex de transmitter
Operating Instructions, 04/2016, A5E02083319-AG
49
7.4
Sensor characteristics
The sensor characteristics menu shows:
●
If a SENSORPROM
®
is installed or not
●
Suppress error P 40 (SENSORPROM
®
not installed)
●
Sensor size
●
Calibration factor
●
Correction factor
●
Excitation
See also sensor characteristics menu diagram (Page 81).
7.5
Reset mode
The reset mode is used for resetting/presetting totalizers or for restoring the factory settings.
Resetting
1.
Press top up key
for 2 sec.
2.
Enter password.
3.
Use forward key
or backward key
to reach reset mode menu.
4.
Press lock/unlock key
to enter reset menu.
5.
Press forward key
to reach totalizer to be reset or default setting menu.
6.
Press lock/unlock key
to start resetting.
If restoring of factory settings (Page 89) is required:
1.
Press lock/unlock key
again to confirm destruction of customized settings.
See also reset menu diagram (Page 82).
7.6
Service mode
All outputs of the transmitter can be forced-controlled in the service mode menu, see also
service mode menu diagram (Page 83).
It is possible to check whether the outputs are functioning.
Error pending and status log lists are also accessible from this menu and the operating time
(in days) can be read.
The forced control is stopped and all previous settings are reinitialized the moment the
service mode is left by pressing top up key
.