Installing/Mounting
4.3 Compact installation
MAG 6000 I / MAG 6000 I Ex de transmitter
24
Operating Instructions, 04/2016, A5E02083319-AG
To ensure optimum flow measurement, attention should be paid to the following:
Vibrations
Image 4-1
Avoid strong vibrations
In installation with strong vibrations remote installation of the transmitter is recommended.
4.3
Compact installation
The flowmeter is delivered with default factory settings and start measuring the flow rate
after power-up. The SENSORPROM
®
is factory-mounted in the transmitter.
Image 4-2
Compact installation
Note
To avoid too high tension on small sizes MAG 1100 sensors, transmitter must be supported.
Summary of Contents for SITRANS F MAG 6000 EX DE