Commissioning
6.3 Changing password
MAG 6000 I / MAG 6000 I Ex de transmitter
42
Operating Instructions, 04/2016, A5E02083319-AG
6.3
Changing password
The setup menu is password-protected in order to ensure that only authorized personnel can
make any changes in transmitter settings.
Change password as follows:
1.
Press top up key
for 2 sec.
2.
Enter password.
3.
Use forward key
or backward key
to reach password menu.
4.
Press lock/unlock key
to unlock password.
5.
Use select key
and change key
to change password.
6.
Press lock/unlock key
to confirm new password.
7.
Press top up key
two times to exit setup mode.
The factory-set password is 1000, but it can be changed to any value between 1000 and
9999.
Factory setting of password can be re-established as follows:
1.
Switch off power supply.
2.
While pressing top up key
- switch on power supply.
3.
Release top up key
after 10 sec.
6.4
Changing basic settings
In the basic settings menu it is possible to set the following parameters:
Parameter
Description
Main frequency
Selection of main power supply frequency corresponding to the country in
which the flowmeter is installed (e.g. 60 Hz in America).
Flow direction
Selection of correct flow direction in pipe.
Customer units
Setting of user defined volume and time units.
Q
max
Setting of measuring range, analog outputs and frequency output. Also indi-
vidual dimension-dependent setting of value, decimal point, unit and time.
Q
max
2
Setting of measuring range, analog outputs and frequency output. Also indi-
vidual dimension-dependent setting of value, decimal point, unit and time. This
menu is only visible if chosen as external digital input.
Summary of Contents for SITRANS F MAG 6000 EX DE