Connecting
5.5 Sensor Installation
FUE1010 IP65 NEMA 4X
80
Operating Instructions, 12/2014, A5E03086491-AC
Actual Zero
The Actual Zero function simply averages the indicated "zero flow" readings (over a user
defined time period) then stores this average value in memory. Under normal operation the
indicated flow reading is zero compensated by simply subtracting this memorized value from
the uncompensated flow reading. Actual Zero is the most positive method for zeroing the
system; however, flow must be stopped with the line blocked (if possible) before invoking this
function. If stopping flow is not possible then an alternate zeroing method should be
selected.
ReversaMatic
This routine involves swapping the Up and Down sensors on the pipe (while keeping the
cables attached) such that the difference in the transit-time change represents the zero
offset. The fixed zero offset value is stored in memory in the same manner as described in
Actual Zero. This routine would generally be used whenever flow cannot be stopped and the
sensors cannot be mounted in the Reflect Mode configuration. Flow must be stable during
the entire process.
ZeroMatic
When ZeroMatic is invoked the flow meter first performs the same analysis as described
above in the AutoZero routine. However, after this analysis is complete the flow meter
continues to interrogate the pipe wall signal and update the zero offset value under normal
operation, such that the flow meter dynamically compensates for changing conditions which
would normally result in zero drift. ZeroMatic will only operate with the sensors mounted in
the Reflect Mode configuration and is recommended for applications which experience large
temperature extremes.
Note
Invoking ZeroMatic will clear any existing "fixed" or memorized zero offset. If any zero offset
remains after flow is stopped, an Actual Zero can be performed without interrupting
ZeroMatic operation. To disable ZeroMatic, invoke it again, but then press <Left Arrow> to
abort the installation.
Note
The ZeroClr command only resets the memorized zero offset registers not those set when
the AutoZero routine is invoked.
Using Actual Zero
Note
Flow must be stopped with the line blocked (if possible) before invoking this function.
Summary of Contents for SITRANS 7ME3500
Page 2: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......
Page 219: ......
Page 220: ......
Page 221: ......
Page 222: ......
Page 223: ......
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 232: ......
Page 233: ......
Page 234: ......
Page 235: ......
Page 236: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 241: ......
Page 242: ......
Page 243: ......
Page 244: ......
Page 245: ......
Page 246: ......
Page 247: ......
Page 248: ......
Page 249: ......
Page 250: ......
Page 251: ......
Page 252: ......
Page 253: ......
Page 254: ...21614 C 1011NFPS 7 ...
Page 255: ...1011NFPS 7 C 21614 ...
Page 256: ...1011NFPS 7 C 21614 ...
Page 257: ...21614 C 1011NFPS 7 ...
Page 258: ......
Page 259: ......
Page 260: ......
Page 261: ......
Page 262: ......
Page 263: ...21614 C ...
Page 264: ......
Page 265: ......
Page 266: ...1012TN 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 267: ...1012TN 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 268: ...1012TNH 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 269: ...1012TNH 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 270: ......
Page 271: ......
Page 272: ......
Page 273: ......
Page 274: ......
Page 275: ......
Page 276: ...21614 C STEP 4 BAND LENGTH TABLE PERMANENT INSTALLATION SERVICE NOTE STEP 2 STEP 3 STEP 1 ...
Page 277: ......
Page 278: ......
Page 279: ......
Page 280: ......
Page 282: ......
Page 290: ...Glossary FUE1010 IP65 NEMA 4X 216 Operating Instructions 12 2014 A5E03086491 AC ...
Page 294: ...Index FUE1010 IP65 NEMA 4X 220 Operating Instructions 12 2014 A5E03086491 AC ...
Page 295: ......