Troubleshooting
10.3 Test Facilities Graph Screen
FUE1010 IP65 NEMA 4X
Operating Instructions, 12/2014, A5E03086491-AC
157
1.
Press the <1> key while viewing the Test Facilities Graph Screen as shown above. The
damping control [MinDamp #] will appear on the command line at the lower left-hand
corner of the screen.
Note
The number listed to the right of the command code on the screen represents the
exponent in the exponential averaging routine (digital damping), where the larger the
number represents the greater the digital averaging. Setting this exponent higher than 7
is generally not recommended.
2.
Pressing the <+> key will increase the MinDamp Factor by one unit for each key press.
To exit this mode, press the <0> key on the keypad.
①
Increased Damping Factor
Figure 10-2 Setting the MinDamp Factor
The above example shows that increasing the Digital Damping reduces asynchronous noise.
To DECREASE the Digital Damping:
Setting the Digital Damping factor to a value LOWER than the default value of 4 may be
justified in cases where pulsating flow is present (such as from a reciprocating pump) or for
the purpose of diagnosing transient signal behavior. A pulsating flow condition that
generates more than +/- 45 degrees of phase jitter will generally cause signal correlation
problems when any digital averaging is used. In this case it may be necessary to completely
eliminate the digital averaging by reducing the Digital Damping Factor to 0.
1.
Press the <2> key while viewing the Test Facilities Graph Screen. The damping control
[MaxDamp #] will appear on the command line at the lower left-hand corner of the screen.
2.
Pressing the <-> key will decrease the MaxDamp Factor by one unit for each key press.
To exit this mode, press the <0> key on the keypad.
Summary of Contents for SITRANS 7ME3500
Page 2: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......
Page 219: ......
Page 220: ......
Page 221: ......
Page 222: ......
Page 223: ......
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 232: ......
Page 233: ......
Page 234: ......
Page 235: ......
Page 236: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 241: ......
Page 242: ......
Page 243: ......
Page 244: ......
Page 245: ......
Page 246: ......
Page 247: ......
Page 248: ......
Page 249: ......
Page 250: ......
Page 251: ......
Page 252: ......
Page 253: ......
Page 254: ...21614 C 1011NFPS 7 ...
Page 255: ...1011NFPS 7 C 21614 ...
Page 256: ...1011NFPS 7 C 21614 ...
Page 257: ...21614 C 1011NFPS 7 ...
Page 258: ......
Page 259: ......
Page 260: ......
Page 261: ......
Page 262: ......
Page 263: ...21614 C ...
Page 264: ......
Page 265: ......
Page 266: ...1012TN 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 267: ...1012TN 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 268: ...1012TNH 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 269: ...1012TNH 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 270: ......
Page 271: ......
Page 272: ......
Page 273: ......
Page 274: ......
Page 275: ......
Page 276: ...21614 C STEP 4 BAND LENGTH TABLE PERMANENT INSTALLATION SERVICE NOTE STEP 2 STEP 3 STEP 1 ...
Page 277: ......
Page 278: ......
Page 279: ......
Page 280: ......
Page 282: ......
Page 290: ...Glossary FUE1010 IP65 NEMA 4X 216 Operating Instructions 12 2014 A5E03086491 AC ...
Page 294: ...Index FUE1010 IP65 NEMA 4X 220 Operating Instructions 12 2014 A5E03086491 AC ...
Page 295: ......