Troubleshooting
10.3 Test Facilities Graph Screen
FUE1010 IP65 NEMA 4X
160
Operating Instructions, 12/2014, A5E03086491-AC
Signal Masking Function: (Hot Key 7)
Under conditions of extremely low signal amplitude, a noise spike associated with the flow
meter receive signal window may be present on the extreme left side of the graph display. If
this spike is large enough it may interfere with the signal detection routines.
1.
To eliminate this noise from the signal processing routines, press the <7> key to invoke
the [Mask Set #] command, then press the <+> key until the noise is no longer present in
the receive waveform.
2.
Press <0> to exit this command.
Hold Set Function: (Hot Key 8)
The [Hold Set #] command is used to set the Hold Set number higher if intermittent mis-
registration occurs. Press the <8> key on the keypad to invoke this function.
Table 10- 3 Hot Key Summary
Key
Command Line
Description
<+>
Expands (magnifies) waveform to view
more detail.
<->
Contracts waveform to view more of
the waveform.
<Left Arrow>
Shifts receive window to the left (wave-
form to the right).
<Right Arrow>
Shifts receive window to the right
(waveform to the left).
<CLR>
Brings waveform to the center of the
screen.
<ENTER>
Calls up Text Display menu items.
<Left Arrow> to return to graph.
<MENU>
Exits the Graph Screen and returns to
the main menu.
<1>
MinDamp
Minimum damping exponent control (+
or - to increase or decrease).
<2>
MaxDamp
Maximum damping exponent control (+
or - to increase or decrease).
<3>
TnSet
Transit time adjustment (use + or - to
move Tn marker).
<4>
ZCOSet
Zero Crossover adjustment (use + or -
to move crossover marker).
<5>
HiSet
Signal envelope threshold level (use +
or - to move threshold). Note: Signal
envelope screen must be activation.
Press = sign to activate
<6>
LoSet
Signal envelope threshold level (use +
or - to move threshold). Note: Signal
envelope screen must be activation.
Press = sign to activate
Summary of Contents for SITRANS 7ME3500
Page 2: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......
Page 219: ......
Page 220: ......
Page 221: ......
Page 222: ......
Page 223: ......
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 232: ......
Page 233: ......
Page 234: ......
Page 235: ......
Page 236: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 241: ......
Page 242: ......
Page 243: ......
Page 244: ......
Page 245: ......
Page 246: ......
Page 247: ......
Page 248: ......
Page 249: ......
Page 250: ......
Page 251: ......
Page 252: ......
Page 253: ......
Page 254: ...21614 C 1011NFPS 7 ...
Page 255: ...1011NFPS 7 C 21614 ...
Page 256: ...1011NFPS 7 C 21614 ...
Page 257: ...21614 C 1011NFPS 7 ...
Page 258: ......
Page 259: ......
Page 260: ......
Page 261: ......
Page 262: ......
Page 263: ...21614 C ...
Page 264: ......
Page 265: ......
Page 266: ...1012TN 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 267: ...1012TN 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 268: ...1012TNH 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 269: ...1012TNH 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 270: ......
Page 271: ......
Page 272: ......
Page 273: ......
Page 274: ......
Page 275: ......
Page 276: ...21614 C STEP 4 BAND LENGTH TABLE PERMANENT INSTALLATION SERVICE NOTE STEP 2 STEP 3 STEP 1 ...
Page 277: ......
Page 278: ......
Page 279: ......
Page 280: ......
Page 282: ......
Page 290: ...Glossary FUE1010 IP65 NEMA 4X 216 Operating Instructions 12 2014 A5E03086491 AC ...
Page 294: ...Index FUE1010 IP65 NEMA 4X 220 Operating Instructions 12 2014 A5E03086491 AC ...
Page 295: ......