SIREMOBIL Compact
SPR2-130.814.01
Page 8 of 10
Siemens AG
System Manual
Rev. 01
03.01
TD PS 24
Medical Solutions
3 - 8
Commissioning
SMPTE test image on the monitor (negative image display)
•
With programmed negative image display, expose a film and measure the density in the
fields 0%, 40% and 100%, enter in Table 2 and compare with the reference values
below.
Table 2: Reference values with programmed negative image display
Gray scale
Minimum
density
Nominal
density
Maximum
density
Actual values
0%
0.16
0.21
0.26
10%
0.20
0.27
0.35
n.a.
20%
0.32
0.42
0.52
n.a.
30%
0.53
0.65
0.78
n.a.
40%
0.85
1.0
1.15
50%
1.28
1.45
1.63
n.a.
60%
1.70
1.90
2.10
n.a.
70%
2.04
2.26
2.49
n.a.
80%
2.31
2.56
2.81
n.a.
90%
2.50
2.77
3.05
n.a.
100%
2.60
2.90
3.20
30%
40%
50%
50%
60%
70%
20%
80%
10%
90%
0%
100%
20%
10%