SIREMOBIL Compact
SPR2-130.814.01
Page 6 of 14
Siemens AG
System Manual
Rev. 01
03.01
TD PS 24
Medical Solutions
2 - 6
Installation
Installation when upgrading in the field
2
Replacing fuse F5
2
Requirements
•
The system must be isolated from the power supply. Unplug the power connector from
the power socket.
•
Open the rear cover of the monitor trolley.
Checking or replacing fuse F5
2
•
Check the value of fuse F5 on board D50 of the ON/OFF assembly (installed in the
bottom shelf of the monitor trolley).
•
If a 4 AT fuse is fitted, replace it with the 6.25 AT fuse supplied.
•
On board D50, stick the label (6.25 AT) supplied on top of the fuse label (4 AT).
•
Manually change the fuse value of the F5 fuse to 6.25 AT in the wiring diagram.
•
Enter the remark "Condonics printer installed" together with the date and your initials.
The system must be isolated from the power supply before begin-
ning work. Unplug the power connector for the system from the
power socket!
WARNING