Installation
2 - 5
Siemens AG
SPR2-130.814.01
Page 5 of 14
SIREMOBIL Compact
Medical Solutions
Rev.01_V 03.01
TD PS 24
System Manual
Wiring the framegrabber
2
•
The power cable is already installed in the monitor trolley and has been connected.
•
Insert the connector from the power cable in the power socket of the framegrabber
(5/Fig. 6).
•
Connect the adapter cable (3.5mm remote jack - BNC connector) from the remote
control to the BNC socket "B&W" and lock (6/Fig. 6).
•
Connect the 3.5 mm jack at the other end of the adapter cable to the 3.5 mm remote
connector (7/Fig. 6). (Remove the heat shrink before)
•
One end of the printer control cable is already connected to the printer. The printer
control cable is fed to the framegrabber via the power chain and has a 25-pin SUB-D
connector at the second cable end.
Connect the 25-pin SUB-D connector to the "PRINTER" socket of the framegrabber
(10/Fig. 6).
Wiring the framegrabber video cable
2
•
Connect one end of the video cable to the 15-pin SUB-D socket of the framegrabber
(8/Fig. 6).
•
With no additional video components present, e.g. a video recorder:
Run the other end of the video cable through the duct (14/Fig. 6) to the MEMOSKOP.
Plug the BNC connector to the "OUT3" socket of the MEMOSKOP and lock.
•
With video recorder present:
Run the other end of the video cable to the video recorder.
•
Connect the BNC connector to the looped-through output of the video recorder and lock.
•
The 75-ohm terminator on the video recorder must be disconnected or unplugged.
Power switch
2
•
Switch on power to the printer (11/Fig. 6).
•
Switch on power to the framegrabber (12/Fig. 6).
•
Load the film cassette with film screens and insert in the printer.
•
Additional steps: see Startup section.
The framegrabber is equipped with a 75-ohm terminating resistor
which cannot be disconnected. If additional video components,
e.g. a video printer or video recorder, are present, the framegrab-
ber must always be wired as the last video component.
NOTICE