Commissioning
3 - 5
Siemens AG
SPR2-130.814.01
Page 5 of 10
SIREMOBIL Compact
Medical Solutions
Rev. 01 03.01
TD PS 24
System Manual
Video frequency and image polarity
•
Press the "Menu" key on the control panel.
•
Select the "Installation" menu line with the cursor key "
↑
" or "
↓
" and press the
"OK/Enter" key.
•
Select the "Select/Setup Host" menu line with the cursor key "
↑
" or "
↓
" on the control
panel of the framegrabber and press the "OK/Enter" key.
•
Select the line "SIEMENS" with the cursor key "
↑
" or "
↓
" and press the "OK/Enter" key.
•
4 selection menus are available:
•
Select the appropriate line with the cursor key "
↑
" or "
↓
" and press the "OK / Enter" key.
Do not perform for 60Hz video frequency and preferred positive
image display (factory default settings)
Menu
Video frequency Image polarity
Description
Siremobil C NTSC
60 Hz or 120 Hz
Positive
Factory setting
Siremobil C PAL
50 Hz or 100 Hz
Positive
for 50 Hz or 100 Hz select
refresh rate and positive image
display
Siremobil NEG 60
60 Hz or 120 Hz
Negative
for 60 Hz or 120 Hz select
refresh rate and negative
image display
Siremobil NEG 50
50 Hz or 100 Hz
Negative
for 50 Hz or 100 Hz select
refresh rate and negative
image display
NOTICE