
Description
3.4 Parameterization
7KT multimeter
Manual, 08/2018, 2514284149-01
35
M09 Alarms (ALAn, n = 1 … 4)
Note
This menu is divided into 4 sections for alarms ALA1 ... 4.
Default
Range
P09.n.01
Alarm source
OFF
•
OFF
•
LIMx
P09.n.02
Channel number (n)
1
1 … 4
P09.n.03
Reset mode
OFF
•
OFF
•
ON
P09.n.04
Priority
Low
•
Low
•
High
P09.n.05
Text
ALAn
(Freely definable text for alarm.
Max. 16 characters)
P09.n.01
Signal which triggers the alarm when a threshold value (LIMx) is exceeded.
P09.n.02
Channel number (n), with reference to the previous parameter.
P09.n.03
•
ON: You must manually save and reset the alarm.
•
OFF: The alarm is saved and reset automatically.
P09.n.04
•
High: In the event of an alarm, the display automatically switches to the
alarm page and displays the alarm icon.
•
Low: In the event of an alarm, the "Information" icon is displayed; the
display does not change.
Summary of Contents for SENTRON 7KT PAC1600 Series
Page 2: ......
Page 10: ......
Page 14: ......
Page 42: ...Installation removal 4 3 Removal 7KT multimeter 40 Manual 08 2018 2514284149 01 4 3 Removal ...
Page 62: ......
Page 69: ...7KT multimeter Manual 08 2018 2514284149 01 67 Dimensional drawings 9 ...
Page 70: ......
Page 75: ......
Page 76: ......