Page 8
D00073611 Siemens SD 270 Service Manual
Socket
Label
Socket Type
Colour
Code
Connected Part
Notes
MIC 1
3.5mm jack
Speech Input Microphone *
MONITOR
3.5mm jack
Monitor Earphone *
BONE
6.3mm jack
Grey
Bone Vibrator Headset *
INSERT
3.5mm jack
Masking Earphone *
RIGHT
LEFT
6.3mm jack
6.3mm jack
Red
Blue
Air conduction headset *
MIC 2
3.5mm jack
Talkback Microphone *
POWER
2.5mm power
jack
Mains AC/DC Adapter *
RESPONSE
6.3mm
jack
Black
Patient Response Switch *
LINE IN
LINE OUT
3.5mm jack
3.5mm jack
CD/tape player
External amplifier
See
2.7
DATA
6 pin mini DIN
Printer *
See
2.5
USB
USB Connector
Computer (via USB port)
See
2.6
The relevant part numbers are indicated in Section 12.
For connected parts marked * only connect the
accessories supplied with the instrument or supplied by
Siemens or a Siemens distributor. These parts have been
tested for use with the Siemens SD 270 Diagnostic
Audiometer for compliance with the standards IEC 60601-
1 and IEC 60601-1-2. The use of accessories other than
those specified may compromise compliance with these
standards. For other sockets refer to Appendix 4.
2.5 Data transfer to a printer
Please refer to Appendix 4 for important information
regarding the connection of non-medical electrical
equipment to medical electrical equipment
The audiometer can be upgraded with an option to allow connection to a
designated portable thermal printer for printing air conduction, bone
conduction, speech and ULL test results (see 3.9). You must use the
designated cable which is supplied with this option.
Upon receipt of the printer it must be initially charged for a minimum of 15
hours prior to use.