P941 Load Cell B, mV
in
Displays the mV signal input from the load cell. Range 0.00 - 60.00 mV.
P942 V/F converter, V
in
Displays the input voltage to the voltage to frequency converter.
Range 0 - 3.98 V
P943 V/F converter, f
out
Displays the output frequency of the voltage to frequency converter.
Range 0 - 131,072
P944 Power sensor
Displays a voltage supply reference for diagnostic purposes.
P949 Diagnostic Errors
Enables or disables diagnostic error checking, E101 - E104
Entry:
0 = disable
1 = enable
Refer to
Troubleshooting
on
page 65
.
P950 Zero register
Registers the number of zero calibrations that have been done since the last master
reset.
P951 Span Register
Registers the number of span calibrations that have been done since the last master
reset.
P952 Design Load
Displays the value of the design load, which corresponds to the full scale value for alarm
and mA output functions. The design load is calculated, based on the design rate and
design speed.
End of test and diagnostic parameters.