Milltronics BW100 - INSTRUCTION MANUAL
Page 33
mmm
mm
Recalibr
ati
on
Initial Zero
Perform an initial zero if necessary when a calibration is out of range message is shown.
Refer to
Maneuvering
on page 18 for parameter selection and value change.
1
2
1
2
1
2
1
2
select P377
invoke initial
zero, enter 1
current zero count
zero calculation
frequency count
displayed
zero deviation
zero deviation accepted
initial zero count = ####
Perform Initial Span