LITHOSTAR Multiline
Medical Solutions
Rev. 02
02.05
CS PS 24
System manual
Final Steps
8
Customer-specific and Country-specific Settings for the Multiline
8
Set requested language according to instruction RXL2-120.032... "Service Software Soft-
ware parameter/monitoring". Parameter Group: / Configuration: User terminal.
Set clinic name according to instruction RXL2-120.032... "Service Software Software
parameter/monitoring. Parameter Group: / Clinic data.
Correct the date and time in the host computer D1 according to instruction "Operation of
the service software" RXL2-120.113...
Back-up and Error Memory
8
•
Perform a back-up and then erase the error memory in the generator and in the Multiline
(refer to Service Software Instructions, RXL2-120.034... and RXL2-120.113...) after
concluding the start-up.
Final Steps
8
•
If the shielding panels have been removed, reinstall them.
•
Attach the labels from register 9 (system binder) at appropriate locations.
•
Install all cover plates, except for the generator front panel and connect the protective
conductors.
•
Measure the protective conductor resistance.
The input field "Actual:" is filled with blanks. These are not visible.
All banks must be deleted before entering the clinic name.
Test meter:
Protective conductor meter
Measurement
value:
max. 0.2 Ohm
(observe country-specific regulations)
(Current 10 A, voltage drop max. 2 V)
Test procedure:
The measurement is made between all conductive parts accessible to
the user and the protective conductor rail in the generator.
Conditions:
Protective conductor resistance: max. 0.2 Ohm
•
If it is greater than this value, increase the wire gauge of the
protective conductor.
•
Reduce the transfer resistance (e.g. tighten the screws).
NOTICE
Unit
Monitor
Cabinets
I.I.
X-ray tube
Collimator
etc.
Prot. conductor meter
≤
0.2 Ohm
Generator