LITHOSTAR Multiline
RXL2-120.815.01
Page 8 of 18
Siemens AG
System manual
Rev. 02
02.05
CS PS 24
Medical Solutions
6 - 8
Checking the Dose Rate
- Reduce or increase the TV iris value (nominal/actual) in the Value Actual field.
If the TV iris value is negative (nominal/actual), it should be reduced.
For example, a value of "- 5" means that the nominal/actual value recorded earlier must
be reduced by - 5 x factor 10, i.e. by - 50.
- Click Set Value after entering the change.
The changed value is marked in the parameter window
(i.e. the value is not yet transferred to the unit).
- Click Put to Unit.
The message "Component... to continue" is displayed.
- Click OK after the automatic Reset.
- Switch off the system for a complete reset.
- Switch on the system.
- Record the dose rate again.
B2. Dose Rate/Iris Adjustment for POLYDOROS SX
6
(Software VE00J)
The dose rate adjustment is performed using the service PC.
The required adjustment menus are described in the Start-up Instructions
"System Configuration with the Service PC", RXL2-120.034.05...
•
Place the 2.1 mm Cu precision filter on the collimator.
•
Center the dosimeter chamber on the I.I.
•
With fluoroscopy ON, check with the smallest zoom format.
•
Select the antiisowatt fluoroscopic curve on the operating PC.
•
Establish a connection between the service PC and the generator.
•
Call up the service software.
•
Enter User Identification and enter User name: Name.
•
Enter Group ID:
_Siemens
.
•
Click OK.
•
Select "System Terminal Setting and Connect: 19200 or 9600" for the Baud rate.
•
Click Login.
•
Confirm the message "Adjusting and/or configuration ..." with OK.
•
Call up POLYDOROS in the Components menu.
•
Switch on S3/X44.
•
Confirm the prompt "Set the S3/X44 switch ..." with OK.
The green LED V90 on D100 lights up.
•
Select Dose rate in the Adjustment menu.
•
If the message "You are ...." appears, confirm with OK.
The message
"
Select II maximum ...
"
appears.
When exiting the dose rate program, switch from the 300 mV
B-signal to the 150 mV B-signal and then back to the 300 mV
B-signal.
NOTICE