Section 7: Troubleshooting, Page 7—5
Cannula openings are
clogged
Remove cannulas from solution
bottles and clean away debris
with cloth and alcohol
Pump tubing is
collapsed from use, or is
perforated
Replace pump tubing. Refer to
Replacement of the Pump
Tubing
.
Underplaten tubing
(from the circuit board
area to the platen) is
clogged with foreign
matter
Replace with new tubing. Refer
to
Yellow
LOW STAIN
light is not functioning
properly
Stain Pak is not properly
positioned in the well
Reposition the Stain Pak so it
moves freely up and down on
the right end
Light is burned out
Replace light assembly. Refer
to
Light switch is not
operating correctly
Contact your local technical
support provider or distributor
for information on adjusting the
switch
Switch or associated
circuit is defective
Contact your local technical
support provider or distributor
Improper mixing action
between stain and
buffer on platen
Instrument is not
properly leveled
Adjust the front feet of the
instrument so the bubble is
centered in the circular bubble
level. Refer to
.
Inadequate pump
volume setting
Contact your local technical
support provider or distributor
if unable to adjust properly.
Uneven staining
Mixing grooves on the
platen are dirty
Clean mixing grooves. Refer to
.
Too much stain or buffer
solution is being
pumped causing poor
mixing of stain and
buffer
Summary of Contents for Hematek
Page 1: ......
Page 6: ...Contents ...
Page 8: ...Page 1 2 Section 1 Introduction ...
Page 22: ...Page 2 2 Section 2 Installation ...
Page 32: ...Page 3 2 Section 3 Operating Instructions ...
Page 38: ...Page 3 8 Section 3 Operating Instructions ...
Page 40: ...Page 4 2 Section 4 Specimens ...
Page 44: ...Page 4 6 Section 4 Specimens ...
Page 46: ...Page 5 2 Section 5 Maintenance ...
Page 54: ...Page 5 10 Section 5 Maintenance ...
Page 56: ...Page 6 2 Section 6 Minor Replacements and Adjustments ...
Page 76: ...Page 7 2 Section 7 Troubleshooting ...
Page 82: ...Page 8 2 Section 8 Service Supplies and Replacement Parts ...
Page 88: ...Page 8 8 Section 8 Service Supplies and Replacement Parts ...
Page 94: ...Page B 2 Appendix B Hematek Preservice Checklist ...
Page 98: ...Index 4 ...