Section 6: Minor Replacements and Adjustments, Page 6—9
5. Remove each tube from the pump assembly.
a. Push the thumb tab on the pump arm to the left as far as possible.
b. While holding the thumb tab in this position, pull the plastic cuff on the tube until
the tube is completely removed from the pump arm.
c. Discard the tube in an appropriate waste container.
Figure 6-10. Removing Tubing from Pump Assembly
6. Remove the new tubes from the Hematek Pump Tube Set.
Each tube is numerically coded to correspond with the numbers shown on the Stain
Pak.
7. Attach the coded end of the new tube to its respective cannula.
a. Replace the cannula if it appears damaged or bent using the Hematek Cannula Set
(PN 4483).
b. If you are re-using any cannulas, make sure the cannula previously used for stain
is used
only
with the stain tubing.
Tube
Code
stain tube
1
buffer tube
2
rinse tube
3
Summary of Contents for Hematek
Page 1: ......
Page 6: ...Contents ...
Page 8: ...Page 1 2 Section 1 Introduction ...
Page 22: ...Page 2 2 Section 2 Installation ...
Page 32: ...Page 3 2 Section 3 Operating Instructions ...
Page 38: ...Page 3 8 Section 3 Operating Instructions ...
Page 40: ...Page 4 2 Section 4 Specimens ...
Page 44: ...Page 4 6 Section 4 Specimens ...
Page 46: ...Page 5 2 Section 5 Maintenance ...
Page 54: ...Page 5 10 Section 5 Maintenance ...
Page 56: ...Page 6 2 Section 6 Minor Replacements and Adjustments ...
Page 76: ...Page 7 2 Section 7 Troubleshooting ...
Page 82: ...Page 8 2 Section 8 Service Supplies and Replacement Parts ...
Page 88: ...Page 8 8 Section 8 Service Supplies and Replacement Parts ...
Page 94: ...Page B 2 Appendix B Hematek Preservice Checklist ...
Page 98: ...Index 4 ...