Page 5—6, Section 5: Maintenance
Emptying the Waste Tank
Empty the waste tank once each day, as well as after installing a new Stain Pak. To avoid
spilling the contents while emptying the waste tank, carefully pull the tank away from the
instrument, supporting the bottom of the tank to hold it level during removal. After
emptying into an appropriate receptacle, rinse the waste tank with water and reinstall it
into the instrument.
Figure 5-3. Removing the Waste Tank
Weekly Cleaning
Clean the drain troughs and the rear guide rail on a weekly basis.
Cleaning Drain Troughs and Rear Guide Rail
The back and front drain troughs of the stainer function as drains for excess stain, and
residues from the reagents that may accumulate and interfere with proper drainage. The
recommended procedure for proper cleaning is as follows:
1. Turn the instrument off and remove the line cord from the AC electrical outlet.
2. Raise the lid of the instrument.
Summary of Contents for Hematek
Page 1: ......
Page 6: ...Contents ...
Page 8: ...Page 1 2 Section 1 Introduction ...
Page 22: ...Page 2 2 Section 2 Installation ...
Page 32: ...Page 3 2 Section 3 Operating Instructions ...
Page 38: ...Page 3 8 Section 3 Operating Instructions ...
Page 40: ...Page 4 2 Section 4 Specimens ...
Page 44: ...Page 4 6 Section 4 Specimens ...
Page 46: ...Page 5 2 Section 5 Maintenance ...
Page 54: ...Page 5 10 Section 5 Maintenance ...
Page 56: ...Page 6 2 Section 6 Minor Replacements and Adjustments ...
Page 76: ...Page 7 2 Section 7 Troubleshooting ...
Page 82: ...Page 8 2 Section 8 Service Supplies and Replacement Parts ...
Page 88: ...Page 8 8 Section 8 Service Supplies and Replacement Parts ...
Page 94: ...Page B 2 Appendix B Hematek Preservice Checklist ...
Page 98: ...Index 4 ...