Page 6—10, Section 6: Minor Replacements and Adjustments
8. Insert tubing into the pump assembly.
a. Extend the new tube to its respective pump and thread the end into the hole in the
pump arm.
b. Push the thumb tab to the extreme left, as before, and push the tube through until
the plastic cuff is flush against the pump arm.
c. Release the pump arm.
NOTE:
If you encounter difficulty
in threading the tubing through the pump housing, lift the
operating lever to
PRIME
for just a few seconds. This will cause the rollers inside the
pump housing to rotate slightly and relieve the interference.
Figure 6-11. Inserting Tubing into Pump Assembly
9. Connect the tube to its proper recessed nipple.
NOTE:
If difficulty is encountered in connecting a tube to its nipple, use forceps or a hemostat to
grasp and attach the tube.
CAUTION
Take care to not damage the tubing when using forceps or hemostats.
10. Replace the cannulas into their respective reagent bottles.
Summary of Contents for Hematek
Page 1: ......
Page 6: ...Contents ...
Page 8: ...Page 1 2 Section 1 Introduction ...
Page 22: ...Page 2 2 Section 2 Installation ...
Page 32: ...Page 3 2 Section 3 Operating Instructions ...
Page 38: ...Page 3 8 Section 3 Operating Instructions ...
Page 40: ...Page 4 2 Section 4 Specimens ...
Page 44: ...Page 4 6 Section 4 Specimens ...
Page 46: ...Page 5 2 Section 5 Maintenance ...
Page 54: ...Page 5 10 Section 5 Maintenance ...
Page 56: ...Page 6 2 Section 6 Minor Replacements and Adjustments ...
Page 76: ...Page 7 2 Section 7 Troubleshooting ...
Page 82: ...Page 8 2 Section 8 Service Supplies and Replacement Parts ...
Page 88: ...Page 8 8 Section 8 Service Supplies and Replacement Parts ...
Page 94: ...Page B 2 Appendix B Hematek Preservice Checklist ...
Page 98: ...Index 4 ...