Section 5: Maintenance, Page 5—5
2. Using the prime function, purge the stain from the tubing.
a. Place the stain cannula, with tubing attached, into a small container of methanol
and lift the operating lever to
PRIME
.
Figure 5-2. Rinsing the Stain Tubing
b. Holding the lever in this position, continue to pump methanol through the stain
tubing until it is thoroughly rinsed and clear solution appears on the platen.
3. After the stain tubing has been cleaned, remove the cannula from the methanol and
continue to prime until the tubing is emptied of all methanol
CAUTION
Always wipe the platen from right to left. Damage to the position or shape of the sensing
switch fingers may result from not following recommended cleaning practices. Check the
position of the fingers after cleaning as described in step 5 of
Instrument Operation Prior to Pump Volume Adjustment
4. Carefully wipe the platen with a soft, disposable, lint-free absorbent cloth or tissue.
Wipe from right to left only.
Summary of Contents for Hematek
Page 1: ......
Page 6: ...Contents ...
Page 8: ...Page 1 2 Section 1 Introduction ...
Page 22: ...Page 2 2 Section 2 Installation ...
Page 32: ...Page 3 2 Section 3 Operating Instructions ...
Page 38: ...Page 3 8 Section 3 Operating Instructions ...
Page 40: ...Page 4 2 Section 4 Specimens ...
Page 44: ...Page 4 6 Section 4 Specimens ...
Page 46: ...Page 5 2 Section 5 Maintenance ...
Page 54: ...Page 5 10 Section 5 Maintenance ...
Page 56: ...Page 6 2 Section 6 Minor Replacements and Adjustments ...
Page 76: ...Page 7 2 Section 7 Troubleshooting ...
Page 82: ...Page 8 2 Section 8 Service Supplies and Replacement Parts ...
Page 88: ...Page 8 8 Section 8 Service Supplies and Replacement Parts ...
Page 94: ...Page B 2 Appendix B Hematek Preservice Checklist ...
Page 98: ...Index 4 ...