9-5
9
Cleaning and Maintenance
Care should be taken not to
scratch the white calibration bar.
Solvents of any kind must not be
used to clean the bar.
Cleaning
Do not push the test table fully
into the analyzer as the test table
may become jammed and prevent
the use of the analyzer.
4
If the white calibration bar is
clean and unmarked, replace
the table into the analyzer by
holding the table at the end
opposite the white calibration
bar, with the white calibration
bar facing upwards. Push the
test table firmly but slowly, just
over half way into the analyzer.
5
Replace the test table insert.
6
If the white calibration bar is
dirty or discolored, gently wipe
and clean it with a new cotton-
tipped stick or lint-free cloth
wetted with distilled water.
7
Allow the calibration bar to air
dry and then inspect the surface
for dust, foreign material,
scratches or scuffs. If the
calibration bar cannot be
cleaned or is still marked, obtain
a new test table.
8
Reinsert the test table as
described earlier in point 4.
Summary of Contents for CLINITEK Status
Page 1: ...Operator s Manual REF 10379682 ...
Page 2: ......
Page 3: ...134798 Rev A 2009 07 Operator s Manual ...
Page 26: ...2 6 ...
Page 28: ...3 2 ...
Page 56: ...4 28 ...
Page 94: ...5 38 ...
Page 98: ...6 4 ...
Page 108: ...7 10 ...
Page 110: ...8 2 ...
Page 132: ...10 14 10 Appendices Appendix E System Overview Principles ...
Page 133: ...10 15 10 Appendices Appendix E System Overview Principles ...
Page 134: ...10 16 10 Appendices Appendix E System Overview Principles ...
Page 135: ...10 17 10 Appendices Appendix E System Overview Principles ...