7-4
7
Troubleshooting
Error Code
Description
Action
E01
Replace the batteries: a) To view instructions
on the display, touch the Error Report selection
area, or b) To use the instructions in this
manual, see page 9 7,
Changing Batteries
.
Low battery power
Lists of Errors and Advisory Messages
List of Errors and Advisory Messages
Clinitek Analyzer:
1. Switch the instrument off by pressing the
on/off button for 2 seconds. 2. Switch the
instrument on again by pressing the on/off
button. 3. Repeat the test.
Loss of test results
E11
1. Make sure that the test table is in place.
Move the test table in or out of the instrument
slightly to reposition the test table. 2. If the
error remains, with the instrument powered
on, unplug the power cord from rear of
instrument and plug back in. Turn instrument
on by pressing the gray power button. 3. If the
error remains with the test table in place,
contact your local representative (contact
information is given in Appendix A).
Failure of test table
E10 or E48
E02
Contact your local representative (contact
information is given in Appendix A).
Failure of
calibration data
E23
Replace the batteries: a) To view instructions
on the display, touch the Error Report
selection area, or b) To use the instructions in
this manual, see page 1 2,
Installing Batteries
and/or page 9 7,
Changing Batteries
. If the
battery level becomes too low to power the
instrument, Error Code E01 will be displayed.
Low battery power
E24
Replace the printer paper a) See instructions
on the inside of the printer paper compartment
cover, or b) To view instructions on the display,
touch the Error Report selection area, or c) To
use the instructions in this manual, see page
1 4,
Loading the Printer Paper or Label Roll.
No printer paper
E12
Contact your local representative (contact
information is given in Appendix A).
Failure of LED
E20
Contact your local representative (contact
information is given in Appendix A).
Failure of clock
Summary of Contents for CLINITEK Status
Page 1: ...Operator s Manual REF 10379682 ...
Page 2: ......
Page 3: ...134798 Rev A 2009 07 Operator s Manual ...
Page 26: ...2 6 ...
Page 28: ...3 2 ...
Page 56: ...4 28 ...
Page 94: ...5 38 ...
Page 98: ...6 4 ...
Page 108: ...7 10 ...
Page 110: ...8 2 ...
Page 132: ...10 14 10 Appendices Appendix E System Overview Principles ...
Page 133: ...10 15 10 Appendices Appendix E System Overview Principles ...
Page 134: ...10 16 10 Appendices Appendix E System Overview Principles ...
Page 135: ...10 17 10 Appendices Appendix E System Overview Principles ...