9-4
9
Cleaning and Maintenance
Cleaning
White
Calibration Bar
BIOHAZARD
Wear personal protective
equipment. Use universal precautions.
Refer to
Appendix H
for recommended
precautions when working with
biohazardous materials.
To enable your Clinitek Status
®
+
analyzer to perform as intended
and provide reliable test results, it
is recommended that you regularly
check the white calibration bar on
the test table, and always check it
after a strip jam.
In normal use, the white calibration
bar should not become dirty or
discolored.
Cleaning the White Calibration Bar
1
Remove the insert from the test
table.
2
Remove the test table by pulling
it slowly out of the analyzer.
3
Check the white calibration bar
on the test table for dirt or
discoloration.