Compact Operating Instructions in accordance with EN 61010-1 and EN 60079-0
40
A5E45779389002-01, 09/2018
Gas inlet conditions
Permissible sample gas pressure
500 to 3000 hPa (7.3 to 43.5 psi) absolute
Purging gas pressure
< 165 hPa (2.4 psi) over ambient pressure
Sample gas flow
18 to 60 l/h (0.3 to 1 l/min)
Sample gas temperature
•
Unheated devices
•
Heated devices
0 to 50 °C (32 to 122 °F)
Max. 15 °C (27 °F) above temperature of analyzer part
Sample gas humidity
< 90% RH (relative humidity), non-condensing
7.4
Technical specifications CALOMAT 6F
Design, enclosure
Weight (analyzer)
Approx. 25 kg (55 lbs.)
Electrical characteristics
Power consumption
Approx. 30 VA
Fuse values
•
100 to 120 V
•
200 to 240 V
1T/250
0.63T/250
Gas path
Gas connections
Stainless steel 1.4571, pipe diameter 1/4" or 6 mm
Gaskets (O-rings)
FFKM (e.g. Kalrez)
Sensor
Silicon, SiO
x
N
y
, gold, epoxy resin, glass
Measuring cell body
Stainless steel 1.4571
Internal gas path (gas lines)
Stainless steel 1.4571
Leak-tightness
Leakage loss < 1
μl/s
Gas inlet conditions
Permissible sample gas pressure
800 to 1100 hPa (11.6 to 16 psi) absolute
Sample gas flow
30 to 90 l/h (0.5 to 1.5 l/min)
Sample gas temperature
0 to 50 °C (32 to 122 °F)
Temperature of measuring cell
60 °C (140 °F)
Sample gas humidity
< 90% RH (relative humidity), non-condensing
Purging gas pressure
< 165 hPa