Technical specifications CALOMAT 62F
Design, enclosure
Weight (analyzer)
Approx. 25 kg (55 lbs.)
Electrical characteristics
Power consumption
•
For non-heated gas connection block
•
For heated gas connection block
Approx. 30 VA
approx. 330 VA
Fuse values
•
For non-heated gas connection
–
100 ... 120 V
–
200 ... 240 V
•
For heated gas connection
–
100 ... 120 V
–
200 ... 240 V
T1L250V (F3/F4)
T630L250V (F3/F4)
T1L250V (F1); T4L250V (F2/F3/F4)
T630L250V (F1); T250L250V (F2/F3/F4)
Gas path
Gas connections
Female thread 1/8"-27 NPT (according to ANSI B 1.20.1)
Gaskets (O-rings)
FPM (e.g. Viton) or FFPM (e.g. Kalrez)
Sensor
Glass
Internal gas path (gas lines)
Stainless steel 1.4571 or Hastelloy C22
Leak-tightness
Loss through leaks < 0.5
μl/s
Gas inlet conditions
Permissible sample gas pressure
800 to 1100 hPa (11.6 to 16 psi) absolute
Sample gas flow
30 to 90 l/h (0.5 to 1.5 l/min)
Sample gas temperature
•
For non-heated gas connection
•
For heated gas connection
0 to 50 °C (32 to 122 °F)
0 to 80 °C (32 to 176 °F)
Temperature of measuring cell
60 °C (140 °F)
Temperature of gas connection with heated gas
inlet
Approx. 70 °C
Sample gas humidity
< 90% RH (relative humidity), non-condensing
Purging gas pressure
< 165 hPa