Compact Operating Instructions in accordance with EN 61010-1 and EN 60079-0
44
A5E45779389002-01, 09/2018
Sample gas inlet conditions
Sample gas inlet conditions of analyzer versions
Type of gas
Non-combustible gases or gas mixtures that
always remain below the LEL in accordance
with Category 4 from EN ISO 13849-1: 2016
Combustible gases up to gas mixtures which
are occasionally explosive
Max. pressure p
max
•
1100 hPa for CALOMAT 6F and
CALOMAT 62F
•
3000 hPa for OXYMAT 6F
•
1500 hPa for ULTRAMAT 6F
•
For combustible gases and analyzer ver-
sions Ex px with "leakage compensation"
–
100 hPa (rel.) for ULTRAMAT 6F and
OXYMAT 6F
•
Analyzer versions Ex px with "continuous
purging" as well as
–
1100 hPa for CALOMAT 6F and
CALOMAT 62F
–
3000 hPa for OXYMAT 6F
–
1500 hPa for ULTRAMAT 6F
Min. pressure p
min
•
500 hPa for OXYMAT 6F
•
600 hPa for ULTRAMAT 6F
•
800 hPa for CALOMAT 6F and
CALOMAT 62F
•
For combustible gases and analyzer ver-
sions Ex px with "leakage compensation"
–
-200 hPa (rel.) for ULTRAMAT 6F and
OXYMAT 6F
Temperature
CALOMAT 6F: 0 to 50 °C
CALOMAT 62F: 0 to 50 °C
OXYMAT 6F: 0 to 50 °C
OXYMAT 6F: In heated version: 0 to 130 °C, but only 15 °C above temperature analyzer unit
ULTRAMAT 6F: 0 to 50 °C
ULTRAMAT 6F: In heated version: 0 to 80 °C
Sample gas flow rate q
v
CALOMAT 6F / CALOMAT 62F: 0.5 ... 1.5 l/min
OXYMAT 6F: 0.3 ... 1.0 l/min
ULTRAMAT 6F: 0.3 ... 1.5 l/min