Medical Solutions
Backup
3.1
NOTE
A backup must be created following software installation, every
software update, and every system adjustment.
NOTE
After a backup group has been successfully burned to CD, the fol-
lowing message is displayed: "Please insert medium in CD-RW in
order to continue."
Confirm this message with "OK".
This does not require a new CD.
•
Select "Backup & Restore" in the Service Software Home Menu.
•
Place the backup CD (located in the monitor cart service compartment) in the CD-ROM
drive.
Fig. 18:
Backup & Restore_Command