Note on the power connection
1.6
The system delivery state is provided by the power connection values on the the line volt-
age label (back of the monitor cart).
Necessary adjustments to local conditions are to be performed as required according to
the adjustment instructions of the ARCADIS Avantic system.
The system leakage current measurement guidelines must be observed.
WARNING
Electrical voltage!
Non-compliance can lead to severe bodily injury and even death.
¹
The internal uninterruptible power supply of the system
(UPS) provides several components with line voltage -
even when the system is switched off or the system
power plug is disconnected.
¹
Before the system is programmed to the local line volt-
age/line frequency, it must be disconnected from the
power supply (the power plug must be disconnected)
and the voltage output plug of the UPS must be discon-
nected as well.