CS PS SP
Specific System Settings
39
Page 39 of 66
Medical Solutions
Fig. 17:
Current limitation
•
Click
<Current limitation>
.
•
Make country-specific changes (only Denmark) and click
<Save>
.
•
Click
<Home>
.
•
Perform a system shutdown.
•
Switch on the system and wait for the system to boot.
NOTE
With the introduction of the Hibernation Mode, system configura-
tion changes are not applied if the system is switched off and then
switched on again using the power buttons on the monitor cart.
To correctly apply the modified data, perform a system shutdown.
Open the <Options> - <End Session> menu and click the “Shut-
down system” button. Immediately thereafter, press the Power off
button on the monitor cart.
The system shuts down and switches off automatically.